Rheonics Calibrator
Calibrator for Type-SR Sensor Probes
- Thermal chamber designed for Rheonics Type-SR Sensor probes (SRV and SRD)
- Integrated solid-state heating and cooling module
- Fluid temperature range: 10 °C (or 15 °C below room temperature) up to 150°C
Specification
15 × 9.5 × 8.5 (inch)
110/220 V AC
max. 360W
11 Kg (24 lb)
10 °C (or 15 °C below room temperature) up to 150°C
max. 65 °C ambient
max. 150 °F ambient
0.005 °C
- SME-BOX (E4)
- Laptop or PC
- RPS
- Class AA , ± (0.10 + 0.0017|T|) °C
How does it work?
Prepare it
The Calibrator uses a Sample Fluid Cylinder that should be filled with the calibration fluid and then place the sensor probe within it.
Assemble it
The Type-SR sensor probe and fluid sample cylinder are placed in the Calibrator.
Start Calibration
The Calibrator Software is guiding the operator to define the fluid parameters, calibration set points, and start the calibration process.
Calibrator
In many industries– including pharmaceuticals, food and beverage, petrochemicals, and chemical processing– regulatory frameworks and quality standards require that measurement instruments carry valid, traceable calibration certificates. Rheonics SRV and SRD sensors are designed for high accuracy and long-term stability. To align with industry standards, the Rheonics Calibrator addresses this topic by verifying the performance of Rheonics SRV inline process Viscometer and SRD Inline process density meter and viscometer under controlled conditions, and certifying the accuracy of the readings.
The Rheonics Calibrator Software performs the measurement, adjusts the calibration if needed, and creates a calibration report, and writes a log onto the sensor electronics (SME).
The Calibrator can be used with any Sensor Module Electronics (SME) variant (SME-TRD,-TR,-DRM, or- BOX (E4)).
Why is the Calibrator useful?
On-Site
Calibration
Calibration
Certificates
High Temperature
Accuracy
Flexibility
Coefficient
Traceability

Calibrator Comparison
Precise and stable temperature control is essential for the optimal operation of Type-SR sensors. The Rheonics TCM line provides highly accurate and consistent temperature regulation of the fluid, ensuring reliable viscosity and density measurements from the start.
Temperature is a critical parameter in fluid behavior, directly influencing particle interactions and, consequently, viscosity. Maintaining a constant and well-controlled temperature is therefore key to achieving accurate and repeatable results.
Compared to conventional laboratory setups—such as thermal baths or ovens with beakers the Rheonics TCM line offers superior temperature stability and control directly within the fluid, minimizing gradients and improving measurement accuracy.
| Rheonics Calibrator vs. Alternative Temperature Control Systems | |||||
|---|---|---|---|---|---|
| Product Characteristics | ![]() | ![]() | ![]() | ![]() | ![]() |
| Name | Rheonics Calibrator | Heat Trace | Viscosity and Density Bath | Laboratory Oven | Climate Chamber |
| Size | 380 x 235 x 220 (mm) | Defined by customer | 420 x 585 x 375 (mm) | 440 x 590 x 465 (mm) | 800 x 800 x 800 (mm) |
| Temperature Range | 10 °C to 150 °C | 65 °C to 200 °C | 5 °C to 100 °C | 30 °C to 250 °C | 10 °C to 350 °C |
| Temperature Accuracy | 0.05 °C | 5 °C | 0.1 °C | 5 °C | 0.05 °C |
Sample Fluid Cylinder
The fluid sample cylinder is made of brass to ensure good temperature uniformity, which is crucial for accurate viscosity and density measurements. Every sample fluid cylinder has a unique serial number. There are three different types and the product key of each is part of the serial number.
Discover more! To see the probe compatibility, please refer to the manual.
Sample Fluid Cylinder with markings
TCC-01
Fluid Volume:
ca. 60 ml
TCC-02
Fluid Volume:
ca. 60 ml
TCC-03
Fluid Volume:
ca. 20 ml
TCM Product line
STCM-IFP
Integrate the device directly in a bypass line for static viscosity analysis. A dedicated valve controls the intake of new fluid samples, while an integrated solid-state module provides active heating and cooling for precise fluid temperature control.
See More
STCM-BB
Take a sample from the process and measure the viscosity and/or density of the fluid. Uses an External bath for temperature control.
STCM-PL
Take a sample from the process and measure viscosity and/or density of the fluid. Control the fluid’s temperature with the in-built heating and cooling module.
See More
DTCM-PL
Ensure uniform temperature distribution across the DVM sensor to enable High-Pressure High-Temperature measurements. Uses a Solid-State Heating and cooling module for temperature control.
DTCM-BB
Ensure uniform temperature distribution across the DVM sensor to enable High-Pressure High-Temperature measurements. Uses an External Bath for temperature control.